A 158-sample study found AI counts close to microscopy, but 77.3% sensitivity at 20% blasts
Study title: Performance Evaluation of an Artificial Intelligence-Assisted Digital Morphology System for Bone Marrow Aspirate Examination
- Evidence type
- Controlled or validated comparison
- Sample
- n = 158
- Sources
- 1 sourcePeer-reviewed
- Published
- 1 min read
- Truvace record
- TRV-2026-1411cite or check it
Study title: Performance Evaluation of an Artificial Intelligence-Assisted Digital Morphology System for Bone Marrow Aspirate Examination
Researchers compared differential cell counts from optical microscopy with those from the CellaVision BMA Application on 158 patient bone marrow aspirate samples. The software was used as intended, preclassifying cells before an expert reviewed them.
Read the study abstract
Introduction: Morphological examination of bone marrow aspirates (BMA) remains essential for the diagnosis of hematologic diseases, but conventional microscopy is labor-intensive and subject to interobserver variability. Methods: A total of 158 samples from patients were analyzed. Differential counts obtained by optical microscopy were compared with those of the CellaVision BMA Application used according to its intended workflow of automated preclassification followed by expert review. The two methods were compa…
How the two sides scoreClose call
The gain
58Gain side
In 158 patient samples, AI preclassification followed by expert review gave cell counts that agreed closely with microscopy, with 96.8% to 99.4% agreement at key decision thresholds.
The problem
57Problem side
Sensitivity at the 20% blast threshold was only 77.3%, and basophil and plasma cell results were weaker, so expert review is still needed.
A fixed formula sets both scores from the evidence and its sources. Reader votes don’t move them.
What the scores mean
G is the gain side, out of 100. It goes up when the gain is specific, when it measurably helped people and when the sources behind it are strong.
P is the problem side, out of 100. It goes up when the problem is specific, when it measurably hurt people and when the sources behind it are strong.
In brief
Counts correlated strongly for most cell types, which suggests this kind of tool could ease a labor-intensive diagnostic task. But weaker results for some cell types and a lower sensitivity at the 20% blast threshold show why human review is still needed, and a single-center series cannot settle wider performance.
Main points
- Pearson correlations ranged from 0.85 to 0.99 for all populations except basophils, which reached 0.68.
- Passing-Bablok regression found no significant systematic or proportional bias for erythroblasts, granulocytic cells, eosinophils and monocytes.
- Bland-Altman mean differences ranged from -1.4% to 1.2% across all populations.
- Plasma cells and basophils showed significant deviations from linearity.
The full story
Bone marrow aspirate morphology remains essential for diagnosing hematologic diseases, yet conventional microscopy is labor-intensive and subject to interobserver variability, which motivated the evaluation of automated preclassification.
Methods included Pearson correlation, Passing-Bablok regression and Bland-Altman analysis, applied overall and within the blast and plasma cell intervals that bracket diagnostic thresholds. A small systematic difference appeared for blasts and a small proportional difference for lymphocytes.
At the clinically relevant blast and plasma cell thresholds, overall agreement between the two methods was high, although the 77.3% sensitivity at the 20% blast threshold marks a gap. The authors describe the results as a basis for further evaluation of this and similar systems.
What this doesn’t fix
The findings come from a single-center series and the system was assessed only within an expert-reviewed workflow, so expert review and further validation remain necessary.
Sources
1 on file- International Journal of Laboratory Hematologydoi:10.1111/ijlh.70263
Talk it over
The AI-assisted counts agreed with microscopy at 96.8% to 99.4% at key thresholds, yet sensitivity at the 20% blast threshold was 77.3%. How do you weigh the broad agreement against that gap?
A 158-sample study found AI counts close to microscopy, but 77.3% sensitivity at 20% blastshttps://truvace.com/item/performance-evaluation-of-an-artificial-intelligence-assisted-digital-morphology
The AI-assisted counts agreed with microscopy at 96.8% to 99.4% at key thresholds, yet sensitivity at the 20% blast threshold was 77.3%. How do you weigh the broad agreement against that gap?
Answer in the threadAll 158 samples came from a single center, and the software was only used with an expert reviewing every result. What would you want to see from other laboratories before deciding the 77.3% sensitivity at the 20% blast threshold holds elsewhere?
Answer in the threadIf you were an expert reviewing the preclassified cells, what would you want to see before trusting the software on basophils and plasma cells, where results were weaker?
Answer in the thread
If you have more time
Conventional microscopy is labor-intensive and subject to interobserver variability. How would you compare the software's errors with the differences between human readers?
Answer in the threadWhat kind of follow-up study, in how many centers, would settle whether the 77.3% sensitivity at the 20% blast threshold is a real limit?
Answer in the thread
ace
What readers say